Pre-use Filter-Integrity Testing: To Test or not to Test?
There is no harmonized guidance on pre-use integrity testing of sterilizing filters, prompting discussion among users as to whether such testing is necessary.


BioPharm International
Volume 25, Issue 8, pp. 16-18

For gamma presterilized single-use systems, where the filter manufacturer has already validated that gamma irradiation does not compromise filter integrity, the risk of filter damage during sterilization appears even more remote. The primary risk of product loss due to integrity failure of gammasterilized filters in a single-use filtration system would most probably be from filter damage incurred after sterilization, for instance, during shipping, handling, and system installation. The lower risk of filter damage and product loss in this case may be further ameliorated if the process is a bulk filtration and, in the event of postuse filter integrity failure, refiltration is qualified and documented in the SOP.

Alternatively, a redundant filtration scheme may be used to minimize risk of product loss in the event of a filter integrity failure discovered only postuse. Finally, the cost of the drug product or process fluid may be sufficiently low that an infrequent disposal is considered inconsequential (e.g., buffers and inexpensive smallmolecule drug products).


Hot Topics at the IBC Single-Use Applications Conference
Although the potential for product loss due to filter nonintegrity may be very low, the effect can be great if the product is expensive and cannot be refiltered, such as with biopharmaceuticals and vaccines. Avoiding catastrophic batch loss, as well as possible product shortages, is a justifiable reason to perform pre-use, poststerilization filterintegrity testing on single-use filtration systems.

Another advantage of single-use systems compared with stainless steel is that versatile system designs can readily accommodate the necessary flushing to prewet filters before the pre-use integrity test. Stainless-steel tanks for the supply and collection of wetting liquid, which must be cleaned and steam sterilized in place, or autoclaved and aseptically connected, can be easily replaced with preassembled flush supply and collection bags that are gamma sterilized in place. Sterilization validation is provided by the supplier. Where product is used to prewet the filter for integrity testing, even the side-flush collection bag can be eliminated in some cases. Additional schemes are also available to enable flushing the filter to reduce leachables and any particles that may be present on the downstream side of the filter, along with wetting them for pre-use integrity testing. Filters can be flushed with product to wet the filter, reducing filter leachables and potential particles while recovering the product to maximize yield.

More information about risk assessment and performance of pre-use integrity testing of gamma-sterilized single-use filtration systems is also available in a webinar led by the author, which is available for free on-demand viewing at http://www.pall.com/biopharmwebinars/.

Jerold Martin is senior vice-president of Global Scientific Affairs at Pall Life Sciences, Port Washington, NY, and chairman of the Board and Technology Committee at Bio-Process Systems Alliance, tel. 516.801.9086,
.

REFERENCES

1. FDA, Sterile Drug Products Produced by Aseptic Processing —Current Good Manufacturing Practice (Rockville, MD, Sept. 2004).

2. EC, Guide to GMP, Annex I — Manufacture of Sterile Medicinal Products (Brussels, Belgium, revised Feb. 2008, effective Mar. 2009).

3. J. Wood, presentation at the IBC Single-Use Conference (San Francisco, CA, Jun. 4–6, 2012).

4 . J. Martin, presentation at the IBC Single-Use Conference (San Francisco, CA, Jun. 4–6, 2012).

5. E. Mahajan et al., Pharmaceutical Engineering, 32 (3), 54–56 (2012).

6 . K. Lear, presentation at the IBC Single-Use Conference (San Francisco, CA, Jun. 4–6, 2012).


blog comments powered by Disqus

ADVERTISEMENT

ADVERTISEMENT

NIH Launches Human Safety Study of Ebola Vaccine Candidate
August 29, 2014
Suppliers Seek to Boost Single-Use Technology
August 21, 2014
Bristol-Myers Squibb and Celgene Collaborate on Immunotherapy and Chemotherapy Combination Regimen
August 20, 2014
FDA Warns about Fraudulent Ebola Treatments
August 15, 2014
USP Awards Analytical Research
August 15, 2014
Author Guidelines
Source: BioPharm International,
Click here