System Suitability. The test system must be properly controlled to ensure reliable release-testing results. The system suitability criteria should
be established during the AMD and optimization phase. This is usually accomplished by running a set of control points. For
each test, system suitability will be satisfied (valid test results generated) if all control points are in established limits.
A test system must be able to reproduce measurable results of a homogeneous sample (control) to allow examination of differences
between batches. Small differences in batches are normal and acceptable, but the sources of variation should be identified.
Ultimately, we will have more certainty when we can separate differences in production batches from assay variability.20
Sample Suitability. Technically, sample suitability is part of system suitability so these parameters can be evaluated together. Sample suitability
should be established during AMD and should ideally ensure that samples, controls, and standards are prepared identically
and run simultaneously. In addition, sample suitability should include a statistical analysis of the number of replicates
needed to generate significant release results. Single measurements may be acceptable if the production-process sampling can
deliver truly batch-representative samples and the precision of assay repeatability is high compared to the product specifications—and
therefore high compared to the batch-to-batch variation on which these specifications are based. Often, however, assay precision
is relatively low, and multiple measurements will substantially increase the level of certainty in the corresponding test
results. Several detailed standard practices are published by ASTM.15–19
Also, we should evaluate batch sampling to ensure capturing and accounting for apparent variability. For example, product
potency in final containers may vary among samples taken from the beginning, middle, and end of fill due to protein adsorption
or protein aggregation during fill. In this case, samples should be taken from each of the three fill stages to ensure a more
accurate representation of the batch's average potency.20
Statistical Data Reduction. Technically, statistical data analysis also is part of overall system suitability. Use multiple statistical values (for instance,
regression line correlation coefficients or p-values for the testing of standard-to-sample line parallelism) to verify that
test system performance is acceptable.
Sometimes, data transformation (for example, logarithmic conversion) may lead to improved linearity. However, most biological
assay response curves are not linear, even after mathematical transformation.20 These are particularly difficult to deal with and should only be handled by experienced statisticians. Nonlinear models,
such as four or five-parameter logistic functions with weighted factors and tests for parallelism, may be the best approach.20 Just as different test methodologies have different biases, changing statistical models may significantly change the final
results. Some models may simply be inappropriate or may not provide acceptable results over the entire assay range.20
Robustness. Robustness, defined as the lack of a significant effect when small changes are deliberately introduced into the test system,
should ideally be addressed during the method optimization phase and not as part of AMV. We should know the degree of robustness
of a method before starting the formal AMV phase. Critical test system parameters (for example, the acceptable range of diluting
the test sample) must be identified and controlled with appropriate operational limits. These limits should be described in
the AMD report and documented in the method SOP. The SOP will then contain operational limits which are in the context of
the overall system suitability criteria and which are adhered to during the validation phase. In addition, robustness should
be tested in the AMD phase during or after method optimization because significant differences in the AMV results (from challenging
the critical operational limits) may be difficult to explain in the AMV report.20
We must remember that AMV is the formal evidence that this method is suitable to be used under strictly controlled QC testing
conditions. The AMV protocol should be set up to deliver this evidence through appropriate acceptance criteria by varying
sample batches and concentrations, operators, instruments, days, and other factors that are expected to vary during routine
testing—in established sample and system suitability conditions and operational limits.20
|