Peak Shape Calibration Method Improves the Mass Accuracy of Mass Spectrometers - - BioPharm International

ADVERTISEMENT

Peak Shape Calibration Method Improves the Mass Accuracy of Mass Spectrometers


BioPharm International
Volume 19, Issue 7

ABSTRACT



The mass accuracy of unit resolution mass spectrometers (MS) is generally near 0.2 Da (~400 ppm at 500 Da) and reported to the nearest integer mass. This is because the software reduces the profile mode spectrum to a centroid spectrum with great loss of information. In this article we will introduce a novel way to calibrate this class of MS instruments that allows routinely achieving 5–10 ppm mass accuracy as well as dramatically improving ion signal extraction and better ratios of signal to noise (S/N).

Centroid Mode Calibration


Equation 1
Three parameters express the mass (σppm) for any mass spectrometer (MS) instrument. It is inversely related to the resolving power (R) of the instrument, the square root of the signal strength (S), and a constant (C) , as shown in this equation,1,2

The constant is determined by factors such as signal unit conversion to real ion counts, peak sampling interval, peak analysis, and mass determination algorithms. The mass accuracy of unit resolution instruments should approach that of high-resolution instruments, but it does not. Instruments based on quadrupole or ion trap technology typically have R values of ~1,000 whereas a high-resolution instrument such as a hybrid quadrupole Time of Flight (qTOFMS) unit has an R value of ~5,000. In theory, the high ion transmitting efficiency (S) of the lower resolution unit should narrow the gap in resolution. In practice, however, the low unit resolution instruments demonstrate a mass accuracy level between 1 and 2 orders of magnitude worse, pointing to an unusually high C value on low-resolution systems.




To understand what impacts the constant, it is first necessary to understand the current approach typically used in acquiring MS data. In most cases, data presented to the user at the end of the acquisition is centroid data, the well-known "stick" graph, in which an ion's m/z (molecular mass/charge) positions are represented by single lines on the graph (Figure 1). This form of graph represents the MS data after heavy processing by the instrument firmware or post-analysis software.

Figure 1 is a flow diagram of a typical instrument data acquisition process. The detector signal is sampled across time (or pixel) to obtain the raw continuum data as output from the detector. A previously established linear or nonlinear calibration equation within the instrument firmware transforms the X-axis to m/z (mass only calibration). This results in what is commonly referred to as a profile mode spectrum.

While most instruments can collect data in the profile mode, it is more common when using unit resolution instruments to let the instrument firmware attempt to locate the peaks from the profile data, and then produce the centroid or stick spectrum before presenting the data to the analyst. This makes for easy data transmission, storage, and management due to the high data rate in a MS system. It made sense in the days of older computer technologies.


blog comments powered by Disqus

ADVERTISEMENT

ADVERTISEMENT

USP Awards Analytical Research
August 15, 2014
FDA Warns about Fraudulent Ebola Treatments
August 15, 2014
Guilty Plea to Importing Illegal Cancer Drugs
August 15, 2014
Amgen Recalls Aranesp Prefilled Syringes
August 15, 2014
Report Predicts Generics Production Return to US
August 15, 2014
Author Guidelines
Source: BioPharm International,
Click here